Anti-Interference: Key Strategies for Precision Weighing in Laboratories

Daicy 2025-07-31 21:41:21

In laboratories, even a microgram-level weighing error can compromise the reliability of experimental results. Environmental interference is the main "invisible killer" behind such deviations. From air flow fluctuations to electromagnetic radiation, from ground vibrations to temperature changes, these seemingly minor factors can cause frequent "distortions" in the readings of precision balances. This article focuses on anti-interference strategies, offering practical advice for laboratory weighing operations to enhance data accuracy.


I. Optimize the Weighing Environment: Minimize Interference at the Source


1.Choose a stable placement location

Place the balance away from equipment that generates air flow or vibration, such as fume hoods, centrifuges, and refrigerators. It is advisable to use a dedicated anti-vibration table or solid concrete platform to avoid interference from ground resonance. Ensure the balance is level by adjusting its bottom feet.


2.Control ambient temperature and humidity

Rapid temperature changes can cause thermal expansion and contraction of internal components, affecting precision. Maintain a constant laboratory temperature (ideally 20±2℃) and use dehumidifiers to keep humidity between 45%-65%, reducing moisture impact on samples or the balance.


3.Avoid electromagnetic interference

Keep the balance away from high-power electrical appliances (e.g., microwaves, autoclaves) and wireless devices like mobile phones or walkie-talkies, as their electromagnetic radiation may disrupt the balance’s circuitry. If necessary, use balances with electromagnetic shielding or add a shielding cover.


II. Standardize Operating Procedures: Reduce Human and Operational Interference


1.Avoid direct contact and air flow disturbance

Wear powder-free gloves during weighing to prevent hand temperature or sweat from affecting sample quality. Avoid rapid body movements or waving to prevent air flow disturbances. For volatile or hygroscopic samples, use sealed containers and minimize lid-opening time.


2.Arrange weighing order and timing rationally

Prioritize high-precision weighing during periods of stable laboratory conditions (e.g., early mornings with fewer operating devices). When weighing multiple samples consecutively, wait for stable readings before proceeding to avoid error accumulation from frequent operations.


3.Regular calibration and maintenance

Even balances with strong anti-interference performance require regular calibration with standard weights, especially after significant temperature changes or equipment relocation. Additionally, clean the weighing pan and sensor regularly to prevent dust or sample residues from reducing sensitivity.


III. Choose Professional Equipment: Enhance Anti-Interference with Technology


In addition to environmental optimization and operational standards, selecting electronic balances with anti-interference designs is crucial for improving weighing reliability. For example, balances with suspended shock-absorbing structures reduce vibration impact; devices with intelligent air flow compensation systems can automatically offset air disturbances; and circuit designs with multi-layer electromagnetic shielding effectively resist external electromagnetic interference.

logo

A leading manufacturer in the weighing industry, committed to precision and innovation in our solutions.

Contact us

Shenzhen Big Dipper Scale Co.,Ltd,No. 88, Tengchang 1st Road, Silian Community,Henggang Sub-district,Longgang District Shenzhen, China

86 0755-28644346

86 13723712960

bdscales@weiaite.com

logo

A leading manufacturer in the weighing industry, committed to precision and innovation in our solutions.

Contact us