Electronic Balance: Fluctuating Readings or Failure to Zero? Step-by-Step Solutions

Daicy 2025-08-30 14:20:56

In daily weighing operations, electronic balances may occasionally experience fluctuating readings or fail to zero after startup. This not only affects weighing accuracy but may also delay work progress. Many users worry about equipment failure when encountering such issues, but in fact, most cases can be resolved through simple troubleshooting. This article will help you quickly restore the normal operation of your electronic balance by covering common causes and step-by-step solutions, while clarifying when professional after-sales support is needed.


I. First, Troubleshoot: These "External Factors" Are the Main Causes 

As precision measuring instruments, electronic balances are sensitive to the operating environment and operational details. In many cases, abnormal readings are not due to equipment failure itself, but to external conditions. We recommend troubleshooting in the following steps first:

 

1. Environmental Factors: Are There "Interference Sources"?

    • Unstable Power Supply: If the balance uses mains power, voltage fluctuations (e.g., when other high-power equipment in the workshop starts) can cause fluctuating readings.
    ✅ Solution: Switch to a regulated power supply (we recommend a regulator with an output voltage error of ±1% or less), or use the balance’s built-in backup battery (ensure the battery is fully charged; replace it promptly if voltage drops below 3.6V). 
    • Airflow/Vibration Interference: Fans, air conditioning vents near the balance, or placement on a vibrating workbench (e.g., near machine tools or conveyors) can cause reading drift.
    ✅ Solution: Move the balance to a wind-free, stable surface and ensure it is level (calibrate using the level bubble at the bottom of the balance; adjust the feet to center the bubble). If environmental vibration is unavoidable, install anti-vibration pads.

Abnormal Temperature/Humidity: Sudden temperature changes (e.g., moving from an air-conditioned room to a high-temperature workshop) or humidity exceeding 65% can affect sensor accuracy and cause fluctuating readings.

✅ Solution: Allow the balance to acclimate to the environment (usually 30 minutes or more) before use. In humid environments, place desiccants around the balance to prevent sensor moisture damage.


2. Operation & Cleaning: Did You Overlook "Detail Issues"? 

    • Abnormal Temperature/Humidity: Sudden temperature changes (e.g., moving from an air-conditioned room to a high-temperature workshop) or humidity exceeding 65% can affect sensor accuracy and cause fluctuating readings.
    ✅ Solution: Allow the balance to acclimate to the environment (usually 30 minutes or more) before use. In humid environments, place desiccants around the balance to prevent sensor moisture damage.
    • Unremoved Blue Protective Film on the Decorative Plate Under the Pan: New balances are shipped with a blue protective film on the decorative plate under the pan (for transportation protection). If parts of the film remain after removal, it may squeeze or block the sensor, causing fluctuating readings.
    ✅ Solution: Turn off the power, remove the pan (follow the manual instructions, usually by lifting gently), and check the decorative plate for residual film. If present, completely remove it with tweezers or a soft cloth. Reinstall the pan and restart the balance after confirming no residue remains.
  • Overloading or Long-Term Non-Calibration: Weighing beyond the balance’s maximum capacity (e.g., weighing 200g on a 100g balance) may damage the sensor. If the balance has not been calibrated for more than 6 months, accuracy drift will cause abnormal readings.
✅ Solution: Check the balance’s capacity label (usually on the side of the body) to confirm no overloading. If calibration is needed, use weights of corresponding accuracy (e.g., Class F1 weights) and perform "manual calibration" according to the manual steps (operations vary slightly by model; refer to the [Product Manuals] section ).


II. Further Check: How to Fix "Internal Minor Issues" of the Equipment 


If the balance still has fluctuating readings or fails to zero after eliminating external factors, check the following equipment-specific details:


1. Loose Connection Cables (for Split-Type Balances) 

Some industrial electronic balances adopt a split design (scale body + indicator). Loose or poor contact of the signal cable connecting the two parts can cause abnormal data transmission.
✅ Solution: Turn off the power, unplug the signal cable (usually DB9 or RJ45 interface), reinsert it firmly and confirm the latch is secured. If the cable is damaged, replace it with original accessories promptly (avoid non-original cables to prevent compatibility issues).

2. Internal Foreign Object Interference: Check for Small Beads, Cockroaches, Etc. 

During use, small objects such as beads, paper scraps, or insects (e.g., small cockroaches) may enter the balance through gaps in the pan. These objects can touch or get stuck on the sensor, causing abnormal sensor force and fluctuating readings.
✅ Solution: Turn off the power and unplug the balance. Use a screwdriver to open the inspection cover on the top (refer to the manual or ask profession engineer to avoid damaging the case). Use a flashlight to check for foreign objects inside. If found, gently remove them with a soft brush or tweezers (do not touch the core part of the sensor). Close the inspection cover after confirming no residue remains, then reconnect the power and restart.

III. Note: Contact Professional After-Sales Support in These Cases 


If the balance still fails to work normally after the above troubleshooting, it may be due to hardware issues such as sensor failure or motherboard damage. Do not disassemble it yourself (to avoid affecting the warranty); contact professional  after-sales team promptly!


logo

A leading manufacturer in the weighing industry, committed to precision and innovation in our solutions.

Contact us

Shenzhen Big Dipper Scale Co.,Ltd,No. 88, Tengchang 1st Road, Silian Community,Henggang Sub-district,Longgang District Shenzhen, China

86 0755-28644346

86 13723712960

bdscales@weiaite.com

logo

A leading manufacturer in the weighing industry, committed to precision and innovation in our solutions.

Contact us